Beam Shaping Devices
Diaphragm Changer
- Positioning accuracy within ±5 µm
- 10B / B4C / Al / Cd or sandwich absorbing plate
- Absorption neutronically tested and certified
- Variable size, variable apertures, custom designed
- Remote controlled
- Easy integration to existing systems
Double Slit Collimation System
- Accuracy within ±10 µm
- 10B / B4C / Al / Cd or sandwich absorbing blades
- Absorption neutronically tested and certified
- Variable size
- Remote controlled
- Easy integration to existing systems
Slit with laser targeting
- Slit with remotely adjustable aperture. X,Y: 0.1-30 mm (customizable)
- Compact, high-resolution industrial camera (5.0 MP)
- Replaceable optics with adjustable focus
- Red Laser Cross Line Module
Monochromator Device
Monochromator goniometric stage
- Rotation around vertical axis
- Double translate (±20 mm)
- Double tilt (±15º)
- High precision positioning accuracy
- Remote controlled
- Movement range customisation
Double focusing monochromator unit
- Horizontal and vertical focusing
- High-resolution positioning accuracy
- Radiation hard components option
- Variable length of crystals up to 300 mm
- Installation and alignment of the
Si crystal sandwich slabs
- Bending mechanism with automatic wedge system
- Remote controlled
Silicon single crystal production
for monochromator
- Double focusing Si bent perfect crystal (BPC) monochromator
- Silicon single crystal slabs asymmetric (311) with (511) option
- Pecision cut for specified orientation
- Orientation measured and certified by X-ray
- Size and stack requirement as per request
- High-precision parallelity by polishing
Multi-Beam Collimation Guide
- Variable size, custom designed
- B4C absorbing inner surfaces
- Absorption neutronically tested and certified
Multi-Positioning Collimation System
- Accuracy within ±10 µm
- Four axis (X, Y, Z, tilting) positioning
- 10B / B4C / Al / Cd or sandwich absorbing blades
- Absorption neutronically tested
- Operation under vacuum or in atmospheric pressure
- Variable size and apertures
(Multi-)Sample Positioning System
- Positional accuracy within ±10 µm
- Heating/Cooling environment
- Magnetic/Non-magnetic environment
- Vacuum option
Collimator production
- Soller Collimators
- Radial Collimators
- Radial Oscillating Collimators
- Optimalization of absorbing element
- Gd2O3 coated PET or textile glass
- Collimator blades transmission are neutronically tested
- Small angle scattering is reduced, certified
- Controlled blade coating thickness: 10 – 100 micron
- Both side of blade could be coated
Shutter production
- Custom designed primary, secondary shutter
- Pneumatic system / stepping motor
- Vertical/horizontal rotating or linear moving
- Optimalization of absorbing element
- Electronic control system
- Single or double beam shutter
- Independent control option for double shutter
- High-precision positioning accuracy
- Customer specified neutron absorption
References »